With the splint fully collapsed and the calibrated Pulley Wheel facing up, position the Sager S300 between the patient’s legs resting the Ischial Perineal Cushion (the saddle) against the Ischial Tuberosity. In the case of a unilateral fracture, the splint should be placed in the Perineum on the side of the injury. In bilateral Fractures excluding Pelvic trauma the side with the greatest degree of injury should be the side of placement. Apply the Abductor Bridle (thigh strap) around the upper thigh of the fractured limb. Tighten the strap snugly. Lift the Spring Clip to extend the Inner Shaft until the Pulley Wheel extends just beyond the Heel. Note that the splint will still perform if an infant is so small that the Wheel extends further.
|
 |
Note the absence or presence of distal pulses, check for sensation. Position the Malleolar Harnesses (ankle harnesses beneath the Heels and just above the Ankles. Fold down the number of Comfort Cushions needed to engage all of the Ankle above the Medial and Lateral Malleoli. Using the attached Hook and Loop Straps, wrap the ankle harness around the Ankle to secure snugly. Note: on very small children with mobile Ankles, it is often necessary to apply tape over the ankle harness and to the skin of the Heel to prevent slippage of the harnesses. Pull the Control Tabs to engage the ankle harness against the Pulley Wheel. This will ensure that the Cable Rings are pulled snugly against the soles of the Feet. Apply QUANTIFIABLE DYNAMIC TRACTION. With one Hand holding the Outer Shaft, gently extend the Inner Shaft of the Splint by pulling it out until the desired amount of traction is recorded on the calibrated Pulley Wheel. It is suggested to use 10% of the patient’s body weight per fractured Femur up to 3 kg (7 pounds) for each Leg. If bilateral fractures are present, the maximum amount would be 7kg (15 pounds), or as directed by the Pediatric Traumatologist. At the hollow of the Knees, gently slide the large elastic Leg Cravat through and upwards to the Thigh, repeating with the smaller Cravats to minimize lower and mid-limb movement. Note: on small infants, one or both of the smaller Cravats may be sufficient for secure immobilization.
|
 |
Adjust the Abductor Bridle (thigh strap) at the upper Thigh making sure it is snug and secure, then firmly secure the elastic Leg Cravats. Apply the Pedal Pinion (figure 8 strap) around the Feet to prevent distal rotation. Note the absence or presence of distal pulses, check for sensation. Patient is now ready for transport.
Warning: All operators should receive full and proper initial/ refresher instruction sessions from a qualified person on detailed use of this equipment and regarding the particular situations in which it should be used.
|
|
Contraindications for traction splints:
All traction splints of any kind are contraindicated in the case of fractured Pelvises unless the Medical Consultant indicates otherwise, or a MAST Trouser has been applied in which case a Sager Splint can be applied over the MAST Trousers. Supracondylar fractures of the Knee and Ankle fractures are also contraindicated. The contraindications listed above are only intended as a basic reference tool. Please defer to federal, state, and/or local protocol for definitive analysis and guidelines.
|
 |
|
Please see our User’s Handbook for detailed Cleaning Instructions.
|
 |
|
Caution: This product contains Natural Rubber Latex, which may cause allergic reactions.
|
|
 |
Recommended traction amount:
How much traction should I apply? Apply the amount of traction recommended by your medical consultant, or that required by protocol. For adults, the American Academy of Orthopedic Surgeons recommends gentle traction to a maximum of 7kg (15 pounds) per fractured Femur [14kg (30 pounds) for a bilateral fracture]. A general rule of thumb is 10% of the patients’ body weight per fractured Femur. For example; if a patient weighing 45kg (100 pounds) has a single fracture, the appropriate amount of traction would be 4 1/2kg (10 pounds). If that same person has a bilateral fracture, 9kg (20 pounds) would be estimated. The Sager S300 Splint is designed to register a maximum of 7kg (15 pounds) of traction. For infants and children, the maximum per fractured Femur would be 3 1/2kg (7 1/2 pounds), 7kg (15 pounds) for bilateral fractures.
|